Test 5.2: Width Jitter
Test Specifications
RMS-<Jitter
0.03% + 25 ps (0.05% + 25 ps for 50 ns
<
width
<
100 ns)
Equipment Needed
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
All cables 50
0,
coaxial, BNC, 122cm (4ft)
High Power Attenuator
Procedure
1. Connect HP 8114A to the Scope as shown:
tip
RI
14A
:IINDEII’
TEST
II
1 BNC CABLE
Equipment Set-up for Jitter
Test
3-20
Testing the HP