EasyManua.ls Logo

HP 8114A - Page 69

HP 8114A
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Take the high-level reading as
Hmz
7. Calculate the Amplitude:
8. Repeat steps 3 to 7 for the following settings:
mble
3-11. Amplitude Levels:
HIZ
(High-Z) into
5OQ
Amplitude Amplitude limit
TR
EntrJ
minimum maximum
2V
1.88
V 2.12 v
6-7
5V
4.85
V
5.15 v 6-8
10 v
9.80 v 10.2 v
6-9
20 v
19.7
v 20.3 v
6-
10
60 v 49.4
v 50.6 V 6- 11
100 v
98.9v 101.1 v
6-
12
Testing the HP 8114A
3-33

Table of Contents