196
Appendix E. Cleaning Procedures
Cleaning Fiber/Panel Connectors
E.3 Cleaning Fiber/Panel Connectors
1 To clean the instrument panel connector remove the connector
interface.
2 Apply some isopropyl alcohol to the lens cleaning paper and
clean the surface and the ferrule of the connectors.
3 Using a new dry piece of cleaning paper, wipe the connector
surface and ferrule until they are dry and clean.
4 Lightly press the adhesive tape several times against the
connector surface to remove any remaining particles. After use
store the tape in the container.
5 Protect the connector surface with a cap.
E.4 Cleaning Connector Interfaces
NOTE If any index matching compound was used, use an ultrasonic bath with
isopropyl alcohol to clean the connector interfaces.
Apply some isopropyl alcohol to the pipe cleaner and wash the
inside the connector interface.
Using a new dry pipe cleaner, dry the inside the connector
interface.
Remove the brush part from the blowbrush and blow air through
the inside the interface to remove any remaining particles.