Model 6485 Picoammeter Instruction Manual DDC Emulation Commands D-5
SRQ
(cont.)
M25 Busy, Reading Done or Reading Overflow
*ESE 1
STATus:MEASurement[:EVENt]:ENABle 192
*SRE <value>
value = (current *SRE value) |= (0x20|0x01)
(B4, B3,
B0)
H
M26 Busy or Reading Done or Data Store Full
*ESE 1
STATus:MEASurement[:EVENt]:ENABle 576
*SRE <value>
value = (current *SRE value) |= (0x20|0x01)
(B4, B3,
B1)
NEW, H
M27 Busy or Reading Done or Data Store Full or Reading
Overflow
*ESE 1
STATus:MEASurement[:EVENt]:ENABle 704
*SRE <value>
value = (current *SRE value) |= (0x20|0x01)
(B4, B3,
B1, B0)
NEW, H
M32 Clears SRQ Error Mask
*ESE 0 (this disables all events in the Standard Event
Register)
*SRE <value>
value = (current *SRE value) &= (~ESB)
(None) X
M33 IDDCO (B5, B0), I
M34 IDDC (B5, B1), I
M35 IDDC or IDDCO (B5, B1,
B0), I
M36 Not in Remote (B5, B2), I
M37 Not in Remote or IDDCO (B5, B2,
B0), I
M38 Not in Remote or IDDC (B5, B2,
B1), I
M39 Not in Remote, IDDC or IDDCO
*ESE 48 Enables the SCPI command and Execution
error events in the standard Event Register
*SRE <value>
value = (current *SRE value) |= (ESB). This Enables an
SRQ on vents enabled in the Standard Event Register.
(B5, B2,
B1, B0), I
Table D-1 (continued)
Device dependent command summary
Mode Command
Description
Equivalent SCPI commands
Note Default