4-6 Keysight B1505A User’s Guide, Edition 12
Installation
Requirements
Installation Requirements
WARNING Do not operate the instrument in dusty environment, or in the presence of
flammable gases, corrosive gases, or fumes.
Ne pas utiliser l’appareil dans un endroit poussiéreux, ou en présence de gaz
inflammables, corrosifs ou de fumée.
Environmental conditions for this instrument are documented in “Operating
Environment” on page 4-5. In principle, this instrument should only be operated
indoors in a controlled environment.
Install the instrument horizontal place then face the roof up.
Fan cools the instrument by drawing air through the sides and exhausting it out
the back. The instrument must be installed in a location that allows sufficient
space at the sides and back of the instrument for adequate air circulation.
If the airflow is restricted, the internal operating temperature will be higher. This
may reduce the instrument’s reliability, or cause the thermal-protection circuits
to turn the instrument off.
Make sure that there is enough space for connecting/disconnecting the power
cable easily.
WARNING Do not put anything on the N1268A Ultra High Voltage Expander.
Ne rien placer sur le N1268A Ultra High Voltage Expander.

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals