EasyManuals Logo

Keysight Technologies B1505A User Manual

Keysight Technologies B1505A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #247 background imageLoading...
Page #247 background image
Keysight B1505A User’s Guide, Edition 12 5-29
Measurement Examples
I-V and Capacitance Measurements Using N1272A
Preparing IV Measurement Setups
This section shows an example of the setup procedure for the Power MOSET IDSS
(drain-source leak current) measurement using the N1272A Selector.
Prepare measurement setups as follows:
1. Click the Application Test tab on the main screen. The tab is in the leftmost
column on the screen.
2. Check the PowerMOSET in the Category field. The Library field lists the test
conditions.
3. Click the IDSS in the Library field and Select button sequentially. The screen
changes the display for setting up the IDSS.
4. Set up the minimum and maximum values and the scale of IDSS at the Device
Parameters field.
5. Set up the measurement conditions at the Test Parameters field.
a. Enter the gate-source bias voltage value in the Gate.
b. Enter or select the drain-source voltage sweep conditions in the Drain.
c. Set up the other parameters if necessary.
For details of the setting parameters, see the help message for each test
definition.

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals