EasyManua.ls Logo

Keysight Technologies B1505A

Keysight Technologies B1505A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight B1505A User’s Guide, Edition 12 5-45
Measurement Examples
Using Multiple HVSMU Modules
Using Multiple HVSMU Modules
This section describes how to perform source output and measurement using the
B1505A installed with multiple HVSMU modules.
“HVSMU Output Range”
“Setting Output Range Mode”
“Performing Measurement
“Available Application Test List”
HVSMU Output Range
If the B1505A is installed with multiple HVSMU modules, specify the output range
mode first to enable the HVSMU. The following lists the available modes.
Single channel mode (initial setting)
An HVSMU channel is available. The other HVSMUs work as the channels of 0
V output with 4 mA compliance.
±1500 V mode
The output range of all HVSMUs is from -1500 V to 0 V and from -8 mA to 0
mA, or from 0 V to +1500 V and from 0 mA to +8 mA.
+3000 V mode
The output range of all HVSMUs is from 0 V to +3000 V and from 0 mA to +4
mA.
-3000 V mode
The output range of all HVSMUs is from -3000 V to 0 V and from -4 mA to 0
mA.

Table of Contents

Other manuals for Keysight Technologies B1505A

Related product manuals