EasyManuals Logo

Keysight Technologies B1505A User Manual

Keysight Technologies B1505A
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
Keysight B1505A User’s Guide, Edition 12 2-47
Introduction
Measurement Resources
MFCMU - Multi Frequency CMU
This section describes typical specification of the multi frequency capacitance
measurement unit (CMU). The CMU performs the impedance measurement and
returns the specified measurement data such as Cp-G. Only one module can be
installed in one mainframe.
Measurement parameters: See Table 2-23.
AC signal, output frequency:
1 kHz to 5 MHz
Setting resolution: 1 mHz (1 kHz to), 10 mHz (10 kHz to), 0.1 Hz (100 kHz to),
or 1 Hz (1 MHz to 5 MHz)
AC signal, output level:
10 mVrms to 250 mVrms, 1 mV resolution
•DC bias:
0 to ± 25 V, using MFCMU internal DC bias
0 to ± 3000 V, using HVSMU and high voltage bias-tee
Measurement range:
For the fixed ranging mode, measurement range (impedance range) must be
specified to perform measurement. Table 2-24 shows the available measurement
ranges and the corresponding impedance value. And Figure 2-11 shows the
calculation example of the impedance vs frequency characteristics of a
capacitive load. Use these information to decide the measurement range.
Impedance Z is calculated by the following formula.
Z = 1 / (2 f C)
where f is frequency (Hz) and C is capacitance (F).

Table of Contents

Other manuals for Keysight Technologies B1505A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B1505A and is the answer not in the manual?

Keysight Technologies B1505A Specifications

General IconGeneral
BrandKeysight Technologies
ModelB1505A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals