Keysight B1505A User’s Guide, Edition 12 2-31
Introduction
Measurement Resources
Table 2-5 HPSMU Measurement Voltage Values and Resolutions
Table 2-6 HPSMU Measurement Current Values and Resolutions
Range
Measurement
Value
a
Measurement Resolutions
High Speed ADC High Resolution ADC
2 V 0  |V| 2.2 V 100 V2 V
20 V 0  |V| 22 V 1 mV 20 V
40 V 0  |V| 44 V 2 mV 40 V
100 V 0  |V| 110 V 5 mV 100 V
200 V 0  |V| 200 V 10 mV 200 V
Range
Measurement Value
a
a. This column is applied to the auto ranging or the limited auto ranging. For
fixed ranging, maximum measurement value is Range column value.
Measurement Resolutions
High Speed ADC High Resolution ADC
1 nA 0  |I| 1.15 nA 50 fA 10 fA
10 nA 0  |I| 11.5 nA 500 fA 10 fA
100 nA 0  |I| 115 nA 5 pA 100 fA
1 A0 |I| 1.15 A 50 pA 1 pA
10 A0  |I| 11.5 A 500 pA 10 pA
100 A0  |I| 115 A 5 nA 100 pA
1 mA 0  |I| 1.15 mA 50 nA 1 nA
10 mA 0  |I| 11.5 mA 500 nA 10 nA
100 mA 0  |I| 115 mA 5 A 100 nA
1 A 0  |I| 1 A 50 A1 A