2-40 Keysight B1505A User’s Guide, Edition 12
Introduction
Measurement Resources
Table 2-15 MCSMU Voltage Ranges and Resolutions
Table 2-16 MCSMU Current Ranges and Resolutions
Range
Output Value and
Measurement Value
Setting/
Measurement
Resolution
Maximum
Current
Maximum
Pulse Base
Value
0.2 V 0 |V| 0.2 V 200 nV
100 mA, 1 A
a
a. For pulse.
± 0.2 V
2 V 0 |V| 2 V 2 V± 2 V
20 V 0 |V| 20 V 20 V± 20 V
40 V 0 |V| 30 V 40 V± 30 V
Range
Output Value and
Measurement Value
Setting/
Measurement
Resolution
Maximum
Voltage
Maximum
Pulse Base
Value
10 A0 |I| 11.5 A 10 pA ± 30 V ± 10 A
100 A0 |I| 115 A 100 pA ± 100 A
1 mA 0 |I| 1.15 mA 1 nA ± 1 mA
10 mA 0 |I| 11.5 mA 10 nA ± 10 mA
100 mA 0 |I| 100 mA,
0 |I| 115 mA
a
a. For pulse.
100 nA ± 100 mA
1 A
b
b. For pulse. Maximum pulse width and duty cycle are 100 ms and 5 %
respectively.
0 |I| 1 A 1 A± 50 mA