2-42 Keysight B1505A User’s Guide, Edition 12
Introduction
Measurement Resources
Table 2-17 HVMCU Voltage Ranges and Resolutions
Table 2-18 HVMCU Current Ranges and Resolutions
Range
Output Value
a
a. Pulse base value is always 0 V.
Measurement
Value
b
b. Measurement is performed by the 3000 V measurement range.
Setting/
Measurement
Resolution
Compliance
Range
Maximum
Current
1500 V 0  |V| 1500 V
0 |V| 2200 V
1.5 mV/3 mV
b
2 A ± 2.5 A
3000 V 0 |V| 2200 V 3 mV/3 mV 1 A ± 1.1 A
100 mA ± 110 mA
Range Output Value
Measurement
Value
Setting
Resolution
Measurement
Resolution
Maximum
Voltage
100 mA 0  |I| 0.11 A 200 nA ± 2200 V
2 A 0 |I| 1.1 A 4 A ± 2200 V
0 |I| 2.5 A 4 A ± 1500 V