Keysight M8070A Programming Guide 101
SCPI Command Reference 5
*CLS
Syntax *CLS
Description This command clears all status register structures in a device. These
registers include:
OPERation Status Register structure
QUEStionable Status Register structure
The corresponding enable registers are unaffected.
*CAL?
Syntax *CAL?
Description This command returns calibration data.
*LRN?
Syntax *LRN?
Description This command gets the device setup query. It return the instrument
settings by binary block data.
*ESE[?]
Syntax *ESE <NRf>
Description The Standard Event Status Enable Command (*ESE) sets the Standard
Event Enable Register. This register acts like a mask, so that the next time
a selected bit goes high, the ESB bit in the status byte is set.
The query (*ESE?) returns the contents of the Standard Event Enable
Register.
*ESR?
Syntax *ESR?
Description This query returns the Standard Event Status Register content. The
register is cleared after it is read.