190 Keysight M8070A Programming Guide
5 SCPI Command Reference
[:SOURce]:INTerference[:LEVel]:CMODe[:STATe][?]
Syntax [:SOURce]:INTerference[:LEVel]:CMODe[:STATe] ‘identifier’, <0|1|ON|OFF>
[:SOURce]:INTerference[:LEVel]:CMODe[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’, ‘M*.DataOut2’, or ‘M*.DataOut’
<0|1|ON|OFF>: Turn common mode interference on/off.
Return Range 0|1
Description This command turns common mode interference on/off.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example :INT:CMOD ‘M2.DataOut1’, on
[:SOURce]:INTerference[:LEVel]:CMODe:AMPLitude[?]
Syntax [:SOURce]:INTerference[:LEVel]:CMODe:AMPLitude ‘identifier’, <NRf>
[:SOURce]:INTerference[:LEVel]:CMODe:AMPLitude? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’ or ‘M*.DataOut2’
<NRf>: Set common mode interference amplitude.
Return Range 0 mV to 320 mV
Description This command sets the amplitude of the common mode interference in
volts. Acceptable units include mV, V and exponents (for example, 100E-3
is the same as 100 mV).
This SCPI is applicable for M8041A and M8051A.
Example :INT:CMOD:AMPL ‘M2.DataOut1’, 100mV

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals