Keysight M8070A Programming Guide 189
SCPI Command Reference 5
[:SOURce]:INTerference[:LEVel]:HFRequency[:SOURce][?]
Syntax [:SOURce]:INTerference[:LEVel]:HFRequency[:SOURce] ‘identifier’, <NRf>
[:SOURce]:INTerference[:LEVel]:HFRequency[:SOURce]? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’ or ‘M*.DataOut2’
<NRf>: Set frequency of HF generator.
Return Range 1 GHz to 6 GHz
Description This command sets the frequency of the high frequency generator.
Acceptable units include MHz, GHz and exponents (for example, 2E9 is the
same as 2 GHz).
This SCPI is applicable for M8041A and M8051A.
Example :INT:HFR ‘M2.DataOut1’, 2.0e9
[:SOURce]:INTerference[:LEVel]:LFRequency[:SOURce][?]
Syntax [:SOURce]:INTerference[:LEVel]:LFRequency[:SOURce] ‘identifier’, <NRf>
[:SOURce]:INTerference[:LEVel]:LFRequency[:SOURce]? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’ or ‘M*.DataOut2’
<NRf>: Set frequency of LF generator.
Return Range 100 Hz to 1 GHz
Description This command sets the frequency of the low frequency generator.
Acceptable units include Hz, kHz, MHz, GHz and exponents (for example,
1E9 is the same as 1 GHz).
This SCPI is applicable for M8041A and M8051A.
Example :INT:LFR ‘M2.DataOut1’, 1.0e9