Keysight M8070A Programming Guide 271
SCPI Command Reference 5
FETCh Subsystem
The FETCh subsystem is used to query the error detector’s results.
Commands for starting, stopping, resetting, etc. are found in the
:PLUGin:ERATio subsystem.
This subsystem has the following SCPI structure:
This subsystem has the following commands:
Table 61
:FETCh:BCOunter?
Syntax :FETCh:BCOunter? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2', or 'M*.DataIn'
Description This query returns the accumulated bit counts and coding specific counts
since start of accumulation period. The results are combined to an
expression. An expression is comparable to a struct in a programming
language. It’s a set of parameter of different types separated be a comma
and enclosed by parenthesis.
The returned expression contains a timestamp, the location and a set of bit
counter values.
(TimeStamp, “Location”, Counted1s, Counted0s, Erroneous1s,
Erroneous0s[, coding specific counters])
Name Description under
:BCOunter? :FETCh:BCOunter? on page 271