258 Keysight M8070A Programming Guide
5 SCPI Command Reference
SENSe Subsystem
The SENSe subsystem represents the analyzer's Data In port.
This subsystem has the following commands:
Table 58
:SENSe:SRATe[?]
Syntax :SENSe:SRATe[?] <identifier>, <NRf>
:SENSe:SRATe? <identifier>
Input
Parameters
‘identifier’: 'M*.DataIn'
Description This command controls the symbol rate for the data input. The symbol rate
can only be controlled directly when the clock source is set to CDR. When
Clk In or Sys Clk is selected as the clock source, then the symbol rate can
only be read.
This query returns the present setting.
This SCPI is applicable for M8046A.
Example :SENS:SRAT ‘M2.DataIn1’,10e9
Name Description under
:SENSe:SRATe[?] :SENSe:SRATe[?] on page 258