224 Keysight M8070A Programming Guide
5 SCPI Command Reference
:OUTPut:EINSertion:DISParity
Syntax :OUTPut:EINSertion:DISParity ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1' or 'M*.DataOut2'
Description This command generates a single disparity error (only valid for 8b/10b
operating mode) on the specified output.
This SCPI is applicable for M8041A, M8051A and M8045A.
Example :OUTP:EINS:DISP ‘M1.DataOut2’