EasyManuals Logo

Keysight Technologies J-BERT M8020A User Manual

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #242 background imageLoading...
Page #242 background image
242 Keysight M8070A Programming Guide
5 SCPI Command Reference
:INPut:COUPling[?]
Syntax :INPut:COUPling ‘identifier’, <AC|DC>
:INPut:COUPling? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<AC|DC>: Select input coupling.
Return Range AC|DC
Description This command selects DC or AC input coupling.
This SCPI is applicable for M8041A and M8051A.
Example :INP:COUP ‘M1.DataIn1’, DC
:INPut:SENSitivity[?]
Syntax :INPut:SENSitivity ‘identifier’, <HIGH|NORMal>
:INPut:SENSitivity? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<HIGH|NORMal>: Select input coupling.
Return Range HIGH|NORM
Description This command sets the error detector input sensitivity to high or normal.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :INP:SENS ‘M1.DataIn1’, HIGH
:INPut[:VOLTage]:WINDow[?]
Syntax :INPut[:VOLTage]:WINDow[?] ‘Identifier’,< NRf>
:INPut[:VOLTage]:WINDow? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn'
Description This command defines the input voltage window. The voltage applied at
the Data In connector needs to be within the voltage range defined by the
common mode voltage and the voltage window.

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals