262 Keysight M8070A Programming Guide
5 SCPI Command Reference
:CLOCk:TRACk[:STATe][?]
Syntax :CLOCk:TRACk[:STATe] <identifier>, <ON|OFF|1|0>
CLOCk:FREQuency? ‘Identifier'
Input
Parameters
‘identifier’: 'M*.DataIn'
Description This command enables or disables the automatic system symbol rate
tracking. When this is enabled, then the system symbol rate is used as the
symbol rate of the addressed functional unit (e.g. DataIn). In this case, the
required clock frequency will be determined automatically based on the
selected frequency multiplier.
See:
• :CLOCk:FREQuency:MULTiplier[?]
• : [:SENSe]:SRATe[?]
This query returns the present state.
This SCPI is applicable for M8046A.
Example :CLOC:TRAC ‘M2.DataIn’,ON