Keysight M8070A Programming Guide 289
SCPI Command Reference 5
:PLUGin:ERATio:ACQuisition:COMPared[?]
Syntax :PLUGin:ERATio:ACQuisition:COMPared ‘identifier’, <NRf>
:PLUGin:ERATio:ACQuisition:COMPared? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: IE+6 to IE+18
Return Range IE+6 to IE+18
Description This command sets the Number of Compared Bits for the addressed
measurement name identifier. The query returns the current value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:ERAT:ACQ:COMP ‘MyMeasurement’, IE+8
:PLUGin:ERATio:ACQuisition:DURation[?]
Syntax :PLUGin:ERATio:ACQuisition:DURation ‘identifier’, <FTIM|IND>
:PLUGin:ERATio:ACQuisition:DURation? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<FTIM|IND>: Set accumulation duration mode.
Return Range FTIM|IND
Description This command sets the accumulation duration mode to either a fixed time
(FTIM) or indefinitely (IND) for the addressed measurement name
identifier. If FTIM is selected, use the :PLUGin:ERATio:ACQuisition:TIME
command to set the duration time. The query returns the current
accumulation duration value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:ERAT:ACQ:DUR ‘MyMeasurement’, FTIM

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals