342 Keysight M8070A Programming Guide
5 SCPI Command Reference
:PLUGin:OLEVel:ACQuisition:ALOCation[?]
Syntax :PLUGin:OLEVel:ACQuisition:ALOCation 'Identifier', < location-string>
:PLUGin:OLEVel:ACQuisition:ALOCation? 'Identifier'
Input
Parameters
Identifier: 'MyMeasurement'
Description This command sets the Location/Location Group string against which the
Data Acquisition is performed. The query returns the current
Location/Location group string against which measurement is configured
for the Data Acquisition.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:OLEV:ACQ:ALOC 'MyMeasurement',"M1.DataIn1"
:PLUGin:OLEVel:ACQuisition:COMPared[?]
Syntax :PLUGin:OLEVel:ACQuisition:COMPared[?] 'Identifier', <NRf>
Input
Parameters
Identifier: 'MyMeasurement'
Return Range 1E+0 to 1E+18
Description This command allows the user to specify the number of compared bits;
that would be used as criteria for moving to the next measurement point.
The query returns current value.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:OLEV:ACQ:COMP 'MyMeasurement', 1E+7

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals