EasyManua.ls Logo

Keysight Technologies J-BERT M8020A - Page 387

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 387
SCPI Command Reference 5
:PLUGin:JTOLerance:BREak
Syntax :PLUGin:JTOLerance:BREak ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to break/pause.
Description This command breaks/pauses a measurement addressed by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example The following example breaks a jitter tolerance measurement with the
measurement name identifier called ‘MyMeasurement’:
:PLUG:JTOL:BRE ‘MyMeasurement’
:PLUGin:JTOLerance:CATalog?
Syntax :PLUGin:JTOLerance:CATalog?
Description This command returns a list of all created jitter tolerance measurement
names currently available for measuring.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example Assume the following is a list of created jitter tolerance measurement
names:
:PLUG:JTOL:NEW ‘JTOL_1’
:PLUG:JTOL:NEW ‘JTOL_2’
:PLUG:JTOL:NEW ‘JTOL_3’
The command and returned list would look like the following:
:PLUG:JTOL:CAT?
“JTOL_1,JTOL_2,JTOL_3”