Keysight M8070A Programming Guide 389
SCPI Command Reference 5
:PLUGin:JTOLerance:DELete
Syntax :PLUGin:JTOLerance:DELete ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to delete.
Description This command deletes a previously created jitter tolerance measurement
addressed by the measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example The following example deletes a jitter tolerance measurement addressed
by the measurement name identifier called ‘MyMeasurement’:
:PLUG:JTOL:DEL ‘MyMeasurement’
:PLUGin:JTOLerance:RESet
Syntax :PLUGin:JTOLerance:RESet ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to reset.
Description This command resets a jitter tolerance measurement addressed by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example The following example resets a jitter tolerance measurement addressed
with the measurement name identifier called ‘MyMeasurement’:
:PLUG:JTOL:RES ‘MyMeasurement’

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals