Appendix E 251
Messages
Error Messages
E. Messages
guide or the company from which you bought the device.
371 Downconverter Unlock local
This error occurs when the local of the downconverter cannot be locked. A device failure is
suspected. Contact Keysight Technologies’ Customer Contact given at the end of this
guide or the company from which you bought the device.
E
-200 Execution error
An error associated with execution has been generated for which this instrument cannot
specify the error message. This code shows the occurrence of an error associated with
execution, as defined in 11.5.1.1.5, IEEE488.2.
F
72 Failed to copy file
This error occurs when copying a file (MMEM:COPY command) fails.
74 Failed to create directory
This error occurs when creating a directory (MMEM:MDIR command) fails.
73 Failed to delete file
This error occurs when deleting a file (MMEM:DEL command) fails.
91 Failed to execute user defined key
In the user menu function, this error occurs when a disabled softkey is executed.
61 Failed to hide trace
In the user window, this error occurs when the show trace “OFF” command is executed to
turn off the remaining trace on the window. At least one trace should be visible at all times.
60 Failed to hide window
This error occurs when the show window “OFF” command is executed to turn off the
remaining window on the screen. At least one window should be visible at all times.
70 Failed to read file
This error occurs when a VBA project file (MMEM:LOAD:PROG command) or other
type of file cannot be read normally.
90 Failed to stop program
This error occurs when stopping a program fails.
71 Failed to write file
This error occurs when the display image (MMEM:STOR:IMAG command) for the LCD
screen, a VBA project file (MMEM:STOR:PROG command) or other type of file cannot
be written normally.
-257 File name error
A file name error. This message appears when an error exists in the file name and thus a
command is not executed correctly. This error occurs, for example, when you try to copy to
an improper file name.