252 Appendix E
Messages
Error Messages
-256 File name not found
The file name specified is not found and thus the command is not executed correctly. This
error occurs, for example, when you try to read a file that does not exist in a disk or a disk
is not correctly inserted into the drive.
80 File transfer failed
This error occurs when writing data into or reading data from a file (MMEM:DATA
command) fails.
I
213 IF A/D overflow
This message appears when the internal IF level is too high and it exceeds the internal A/D
converter input range. In the phase noise measurement, check that the DUT’s carrier signal
is within the selected frequency band. Also, the message may appear if a DUT that has a
large noise level and an unstable output level is measured. When measuring phase noise in
such a case, decrease the IF Gain value or increase the sweep start value if possible. In the
spectrum monitor measurement, decrease the DUT’s carrier level, or set the reference level
or input attenuator value larger than the current setting. In the transient measurement,
decrease the DUT’s carrier level, or set the max input level or input attenuator value larger
than the current setting.
260 IF Level Overload
The IF level is too high. Set the input attenuator value larger than the current setting.
312 IF not found
The IF signal cannot be found. In the phase noise measurement, this error occurs when no
signal is inputted to the RF IN, or the input level to the RF IN is too low. Also, the message
may appear if a DUT that has an unstable output level is measured, or if the carrier
frequency of DUT is out of the specified frequency band. Aside from the above reasons,
there is the possibility of a device failure. Contact a Keysight Technologies sales office or
the company from which you bought the device.
-224 Illegal parameter value
The parameter value is improperly set.
-282 Illegal program name
This error occurs when a nonexistent VBA program name is specified by the
PROG:SEL:NAME command.
-213 Init ignored
Because another measurement is in progress, the request for initiating a measurement
(“INIT” command) is ignored.
250 Insufficient IF Level
The IF level is too low. Set the input attenuator value smaller than the current setting.
230 Insufficient RF Level
The input level to the RF IN connector is too low.
50 Internal test failed
Internal test failed.