254 Appendix E
Messages
Error Messages
See the command reference to confirm the required number of parameters.
200 Power on test failed
This error occurs when the power-on test fails, indicating a failure of the instrument.
Contact a Keysight Technologies sales office or the company from which you bought the
instrument.
40 Printer error
This error occurs when the previous printing is still in progress or the printer fails (offline,
short of paper, etc.) at the time of outputting the display image on the LCD screen to the
printer (HCOP:IMM command).
41 Print failed
This error occurs when printing fails for reasons other than Error 40, Printer error.
-284 Program currently running
This error occurs when the PROG:SEL:STAT RUN command is executed when the VBA
program is in the Run state.
-286 Program runtime error
An error occurring when VBA is executed.
R
75 Recall failed
This error occurs when reading an instrument status file (State01.sta, etc.)
(MMEM:LOAD:STAT command) fails.
220 RF freq out of range
This error occurs when the DUT’s output frequency is not within the measurement range.
240 RF level overload
This error occurs when the input to the RF IN port exceeds the maximum input level in the
measurement. The measurement value obtained in such a case is not correct. In the worst
case, a failure (damage to the receiver) may occur.
S
76 Save failed
This error occurs when writing an instrument status file (State01.sta, etc.)
(MMEM:STOR:STAT command) fails.
78 Save VBA program failed
This error occurs when saving a VBA program file fails.
210 System PLL unlocked
This error occurs when the PLL circuit of the instrument becomes unlocked while the
measurement is in progress. This message may appear if the DUT’s carrier signal is
unstable. If this message is displayed in the phase noise measurement, the following
problems and corrective actions are shown below.
The DUT’s noise level is too large, or a large level spurious component exists in the