Each power module has a fully integrated voltmeter and ammeter to measure the actual voltage and
current being sourced from the output to the device under test. The accuracy of the voltage and
current measurements is based on the type of power modules that are installed, as documented in
the Keysight N6700 DC Power Analyzer Family Specifications Guide .
Meter View
Each output has its own measurement capability. When the Meter view is displayed, the
measurement system continuously measures the output voltage and current. The measurement
system acquires as many points as needed based on the number of power line cycles (NPLC) and the
time interval, and averages the samples.
As shown below, the default view on the left displays all four outputs. The single-output view on the
right displays more information about the selected output. Press the Meter View key to toggle
between the two views.
From the remote interface:
To trigger and return DC current or voltage measurements:
MEAS:CURR? (@1)
MEAS:VOLT? (@1)
To return a previously triggered DC current or voltage:
FETC:CURR? (@1)
FETC:VOLT? (@1)