The current or voltage value of ten steps:
ARB:CURR:CDW 0.5,1,1.5,2,2.5,3,3.5,4,4.5,5,(@1)
ARB:VOLT:CDW 1,2,3,4,5,6,7,8,9,10,(@1)
Exponential Parameters
Parameter Description
Start Setting (I
0
or V
0
) The setting before the waveform.
End Setting (I
1
or V
1
) The end setting of the waveform.
Delay (T
0
) The delay after the trigger is received but before the waveform starts.
Time (T
1
) Time for the amplitude to go from the start setting to the end setting.
Time Constant(T
c
) The time constant of the curve.
From the remote interface:
The setting before the waveform:
ARB:CURR:EXP:STAR 0,(@1)
ARB:VOLT:EXP:STAR 0,(@1)
The end setting of the waveform:
ARB:CURR:EXP:END 1,(@1)
ARB:VOLT:EXP:END 5,(@1)
The delay after the trigger is received but before the waveform starts:
ARB:CURR:EXP:STAR:TIM 0.25,(@1)
ARB:VOLT:EXP:STAR:TIM 0.25,(@1)
Time for the amplitude to go from the start setting to the end setting:
ARB:CURR:EXP:TIM 0.75,(@1)
ARB:VOLT:EXP:TIM 0.75,(@1)
The time constant of the curve:
Keysight N6705C Operating and Service Guide 111