EasyManua.ls Logo

Keysight N6705C - Exponential Parameters

Keysight N6705C
476 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The current or voltage value of ten steps:
ARB:CURR:CDW 0.5,1,1.5,2,2.5,3,3.5,4,4.5,5,(@1)
ARB:VOLT:CDW 1,2,3,4,5,6,7,8,9,10,(@1)
Exponential Parameters
Parameter Description
Start Setting (I
0
or V
0
) The setting before the waveform.
End Setting (I
1
or V
1
) The end setting of the waveform.
Delay (T
0
) The delay after the trigger is received but before the waveform starts.
Time (T
1
) Time for the amplitude to go from the start setting to the end setting.
Time Constant(T
c
) The time constant of the curve.
From the remote interface:
The setting before the waveform:
ARB:CURR:EXP:STAR 0,(@1)
ARB:VOLT:EXP:STAR 0,(@1)
The end setting of the waveform:
ARB:CURR:EXP:END 1,(@1)
ARB:VOLT:EXP:END 5,(@1)
The delay after the trigger is received but before the waveform starts:
ARB:CURR:EXP:STAR:TIM 0.25,(@1)
ARB:VOLT:EXP:STAR:TIM 0.25,(@1)
Time for the amplitude to go from the start setting to the end setting:
ARB:CURR:EXP:TIM 0.75,(@1)
ARB:VOLT:EXP:TIM 0.75,(@1)
The time constant of the curve:
Keysight N6705C Operating and Service Guide 111
3 Using the Source Functions

Table of Contents