Ramp Parameters
Parameter Description
Start Setting (I
0
or V
0
) The setting before the ramp.
End Setting (I
1
or V
1
) The setting after the ramp.
Delay (T
0
) The delay after the trigger is received but before the ramp starts.
Ramp Time (T
1
) The time that the output ramps up.
End Time (T
2
) The time the outputremains at the end setting after the ramp completes.
From the remote interface:
The setting before the ramp:
ARB:CURR:RAMP:STAR 0,(@1)
ARB:VOLT:RAMP:STAR 0,(@1)
The setting after the ramp:
ARB:CURR:RAMP:END 1,(@1)
ARB:VOLT:RAMP:END 5,(@1)
The delay after the trigger is received but before the ramp starts:
ARB:CURR:RAMP:STAR:TIM 0.25,(@1)
ARB:VOLT:RAMP:STAR:TIM 0.25,(@1)
The time that the output ramps up:
ARB:CURR:RAMP:RTIM 0.5,(@1)
ARB:VOLT:RAMP:RTIM 0.5,(@1)
The time the output remains at the end setting after the ramp:
ARB:CURR:RAMP:END:TIM 0.01,(@1)
ARB:VOLT:RAMP:END:TIM 0.01,(@1)