Step Parameters
Parameter Description
Start Setting (I
0
or V
0
) The setting before the step.
End Setting (I
1
or V
1
) The setting after the step.
Delay (T
0
) The delay after the trigger is received but before the step occurs.
From the remote interface:
The setting before the step:
ARB:CURR:STEP:STAR 0,(@1)
ARB:VOLT:STEP:STAR 0,(@1)
The setting after the step:
ARB:CURR:STEP:END 1,(@1)
ARB:VOLT:STEP:END 5,(@1)
The delay after the trigger is received before the step occurs:
ARB:CURR:STEP:STAR:TIM 0.01,(@1)
ARB:VOLT:STEP:STAR:TIM 0.01,(@1)
Trapezoid Parameters
3 Using the Source Functions
116 Keysight N6705C Operating and Service Guide