Parameter Description
Start Setting (I
0
or V
0
) The setting before the staircase.
End Setting (I
1
or V
1
) The setting after the final stair step. Notethatthe differencebetween the startand end setting is divided
equally between the steps.
Delay (T
0
) The delay after the trigger is received but before the staircase starts.
Step Time (T
1
) The time to completeall staircase steps.
End Time (T
2
) The time the outputremains at the end setting after the staircasecompletes.
# of Steps The total number of staircase steps.
From the remote interface:
The setting before the staircase:
ARB:CURR:STA:STAR 0,(@1)
ARB:VOLT:STA:STAR 0,(@1)
The setting after the final stair step:
ARB:CURR:STA:END 1,(@1)
ARB:VOLT:STA:END 5,(@1)
The delay after the trigger is received but before the staircase starts:
ARB:CURR:STA:STAR:TIM 0.2,(@1)
ARB:VOLT:STA:STAR:TIM 0.2,(@1)
The time to complete all staircase steps:
ARB:CURR:TIM 0.2,(@1)
ARB:VOLT:TIM 0.2,(@1)
The time the output remains at the end setting after the staircase:
ARB:CURR:STA:END:TIM 0.2,(@1)
ARB:VOLT:STA:END:TIM 0.2,(@1)
The total number of staircase steps:
ARB:CURR:STA:NST 3,(@1)
ARB:VOLT:STA:NST 3,(@1)
Keysight N6705C Operating and Service Guide 115