EasyManua.ls Logo

Keysight N6705C - Page 115

Keysight N6705C
476 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter Description
Start Setting (I
0
or V
0
) The setting before the staircase.
End Setting (I
1
or V
1
) The setting after the final stair step. Notethatthe differencebetween the startand end setting is divided
equally between the steps.
Delay (T
0
) The delay after the trigger is received but before the staircase starts.
Step Time (T
1
) The time to completeall staircase steps.
End Time (T
2
) The time the outputremains at the end setting after the staircasecompletes.
# of Steps The total number of staircase steps.
From the remote interface:
The setting before the staircase:
ARB:CURR:STA:STAR 0,(@1)
ARB:VOLT:STA:STAR 0,(@1)
The setting after the final stair step:
ARB:CURR:STA:END 1,(@1)
ARB:VOLT:STA:END 5,(@1)
The delay after the trigger is received but before the staircase starts:
ARB:CURR:STA:STAR:TIM 0.2,(@1)
ARB:VOLT:STA:STAR:TIM 0.2,(@1)
The time to complete all staircase steps:
ARB:CURR:TIM 0.2,(@1)
ARB:VOLT:TIM 0.2,(@1)
The time the output remains at the end setting after the staircase:
ARB:CURR:STA:END:TIM 0.2,(@1)
ARB:VOLT:STA:END:TIM 0.2,(@1)
The total number of staircase steps:
ARB:CURR:STA:NST 3,(@1)
ARB:VOLT:STA:NST 3,(@1)
Keysight N6705C Operating and Service Guide 115