New Feature Comparison With A and B Power Analyzers
Feature N6705C Implementation N6705A/B Implementation
Color Phantom gray Quartz gray
Start-up time Approximately 30 seconds Approximately 15 seconds
Compliance LXI 1.4 Core, HiSLIP, IPv6 LXI 1.1 Class C
Ethernetconnectivity Up to 1000 Mbps Up to 100 Mbps
Auto-MDIX Supported Not available
*IDN response Keysight Technologies
N6705C
Agilent Technologies
N6705A, N6705B
Firmwareupdatemethod FirmwareUpdateUtility only Firmware Update Utility
or front panel USB port
Firmwareupdatelockout from
front panel
Requires Admin password to update firmware Not available
ID strings - LAN, USB, GPIB,
and Web page
Keysight, K-<hostname> Agilent, A-<hostname>
Low battery detectatstart-up Supported Not available
Saved state locations 10 2
Web control No Java plug-in required Java plug-in required
Calibration count CALibrate:COUNt Returns -100 Command error
Display image capturecom-
mands
HCOPy:SDUMp:DATA?
HCOPy:SDUMp:DATA:FORMat
Returns -100 Command error
Front panel display enable/dis-
able
DISPlay[:WINDow][:STAte] Returns -100 Command error
LXI identify and MDNS com-
mands
LXI:IDENtify[:STATe]
LXI:MDNS[:STATe]
Returns -100 Command error
Differencebetween the internal
temperature and the OT trip
point
OUTPut:PROTection:TEMPerature:MARgin Returns -100 Command error
PERSona commands switch
manufacturer and model num-
ber for backward compatibility
SYSTem:PERSona:MANufacturer
SYSTem:PERSona:MANufacturer:DEFault
SYSTem:PERSona:MODel
SYSTem:PERSona:MODel:DEFault
Returns -100 Command error
Memory sanitization NISPOM
compatible
SYSTem:SECurity:IMMediate Returns -100 Command error
Set the instrument statein bin-
ary format
SYSTem:SET Returns -100 Command error
7 SCPI Programming Reference
370 Keysight N6705C Operating and Service Guide