Performance Verification
Verification Description
Verification Test Setups
Constant Voltage Tests
Transient Recovery Time
Auxiliary Voltage Measurement Accuracy (N6781A and N6785A only)
Constant Current Tests
Resistance Programming Accuracy (N6781A and N6785A only)
Negative Current Output (N6783A-BAT and -MFG only)
Verification Description
The verification procedures described in this section verify that the Keysight N6705C DC Power
Analyzer is operating normally and is within published specifications. The following types of
performance verification tests are available:
Performance Tests - These tests verify that the power analyzer is operating normally and meets the
specifications published in the Keysight N6700 DC Power Analyzer Family Specifications Guide.
Calibration Tests - These tests verify that the power analyzer is operating within its calibration limits.
Keysight Technologies recommends that complete verification should always be performed at the
calibration interval. If the power analyzer passes the verification tests, the unit is operating within its
calibration limits and does not need to be re-calibrated. This ensures that the instrument will remain
within specifications for the next calibration interval and provides the best long-term stability.
Performance data measured using this method may be used to extend future calibration intervals.
If the instrument fails any of the tests or if abnormal test results are obtained, try calibrating the unit. If
calibration is unsuccessful, return the unit to a Keysight Technologies Service Center.
8 Verification and Calibration
374 Keysight N6705C Operating and Service Guide