Emulation Command
[SOURce:]EMULation <type>, (@<chanlist>)
[SOURce:]EMULation? (@<chanlist>)
Specifies the emulation type for models N678xA. Refer to Emulation Mode Settings for information on
the default parameter settings for each of the following emulation modes:
PS4Q Four-quadrant bipolar power supply (model N6784A only). Power supply operation is availablein all four output quadrants with
no restrictions.
PS2Q Two-quadrant bipolar power supply. In this mode operation is restricted to two quadrants (+V/+I and +V/- I).
PS1Q Unipolar, one-quadrant power supply. This mode emulates a typical one quadrant or unipolar power supply with limited down-
programming.
BATTery Two-quadrant battery emulator. A battery emulator imitates a battery’s charging and discharging functions.
CHARger One-quadrant battery charger. A battery charger imitates a battery charger; itcannot sink current like a battery.
CCLoad One-quadrantload. This mode emulates a constant-currentload.
CVLoad One-quadrantload. This mode emulates a constant-voltage load.
VMETer Voltage measurement. Thevoltagemeasurement emulates a voltmeter.
AMETer Current measurement. The current measurement emulates a zero-burden ammeter.
Parameter Typical Return
<type>: PS4Q | PS2Q | PS1Q | BATTery | CHARger |
CCLoad | CVLoad | VMETer | AMETer,
*RST PS4Q for N6784A; PS2Q for N6781A, N6782A,
N6785A, N6786A
PS4Q, PS2Q, PS1Q, BATT, CHAR, CCL, CVL,
VMET, AMET
Selects CC load as the emulation type for channel 1: EMUL CCL, (@1)
7 SCPI Programming Reference
280 Keysight N6705C Operating and Service Guide