Power Model Differences
Feature
(l = available)
DC Power High-Performance Precision
N673xB, '4xB, '7xA N675xA N676xA
50 W outputrating N6731B – N6736B N6751A N6761A
100 W output rating N6741B – N6746B N6752A N6762A
300 W output rating N6773A – N6777A N6753A,'54A N6763A, '64A
500 W output rating N6755A, '56A N6765A, '66A
Outputdisconnect relays Option 761 Option 761 Option 761
Outputdisconnect/polarity reversal relays
1
Option 760 Option 760 Option 760
Arbitrary waveform generation
l l l
Autoranging outputcapability
l l
Voltage/current turn-on priority N6761A, '62A
Precision voltage/current measurements
l
Low voltage/current outputrange N6761A, '62A
Low voltage/current measurement range
l
200 microamperemeasurement range
2
Option 2UA
Voltage/current scopetraces
l l l
Simultaneous voltage/current scope traces
l
Simultaneous voltage/current datalogging
3
l
Interleaved voltage/currentdata logging
3
l l
Dynamic current correction
l
N6751A, '52A N6761A, '62A
SCPI command output list capability
4
l l l
SCPI command array readback
4
l l l
SCPI command programmable sample rate
4
l l l
SCPI command external data logging
4
l l l
Double-wide power modules
(require 2 channel locations)
N6753A - N6756A N6763A - N6766A
Note 1 Output current is limited to 10A max. on Models N6742B and N6773A with Option 760.
Option 760 is not available on Models N6741B, N6751A, N6752A, N6761A, and N6762A.
Note 2 Option 2UA.is only available on Models N6761A and N6762A. It includes Option 761.
Note 3 Option 055 deletes the Data Logger function on Model N6705C.
Note 4 Only available when using the remote interfaces; not from the front panel.
Keysight N6705C Operating and Service Guide 41
1 Quick Reference