N678xA Power Model Differences
Feature
(l = available)
Source/Measure Units (SMU) Application-Specific
N6781A N6782A N6784A N6785A N6786A N6783A
-BAT
N6783A
-MFG
Outputrating 20 W 20 W 20 W 80 W 80 W 24 W 18 W
2-quadrant operation
l l l l l l
4-quadrant operation
l
Aux voltage measurement
l l
Outputdisconnect relays
l l l l l
Opt 761 Opt 761
Arbitrary waveforms
1
l l l l l l l
Negativevoltage protection
l l l l l l l
Voltage or currentpriority
l l l l l
CC load or CV load
l l l l l
Voltage/current measurement only
l l l l l
Battery emulator or charger
l l
Programmable resistance
l l
Voltage output ranges 3 3 3 4 4 1 1
Current outputranges 3 3 4 4 4 1 1
Voltage measurement ranges 3 3 3 1 1 1 1
Current measurement ranges 4 4 4 3 3 2 2
Voltage/current scopetraces
l l l l l l l
Simultaneous voltage/current measurements
l l l l l
Simultaneous voltage/current datalogging
2
l l l l l
Interleaved voltage/currentdata logging
2
l l
Seamless measurement autoranging
l l l l
SCPI output lists
1,3
l l l l l l l
SCPI array readback
3
l l l l l l l
SCPI programmable sample rate
3
l l l l l l l
SCPI external data logging
3
l l l l l l l
SCPI histogram measurements
3
l l l l
Double-wide power modules
l l
Note 1 Arbitrary waveforms and List capability are not available on the negative current output on Model N6783A.
Note 2 Option 055 deletes the Data Logger function on Model N6705.
Note 3 Only available when using the remote interfaces; not the front panel.
1 Quick Reference
42 Keysight N6705C Operating and Service Guide