SENS:DLOG:OFFS 25
To trigger the Data Logger measurement:
TRIG:DLOG (@1)
(If the trigger source is BUS, you can also program *TRG or <GET>.)
Data Logger Filename
From the front panel:
Select the Filename button to specify a filename in which to save the datalog. Data will be logged to
this filename the next time the Data Logger runs. If you do not specify a filename, the data will be
logged to default.dlog, which is overwritten each time the Data Logger runs.
Enter the filename in the Path\File field. Check Append date and time at start of log to include time-
stamp information in the file.
From the remote interface:
To specify the internal filename in which to save the datalog:
INIT:DLOG datalog1.dlog”
You can also export the datalog that is presently displayed in the data logger view. To export the
datalog to the specified filename, send:
MMEM:EXP:DLOG datalog1.csv”
To export a datalog that has been saved in the instrument, you must first load the saved
file into the data logger view.