List commands program an output sequence of multiple voltage or current settings. A comma-
delimited list of up to 512 steps may be programmed. Note that these commands only apply in the
presently active priority mode, either voltage priority or current priority.
[SOURce:]LIST:COUNt <value>, (@<chanlist>)
[SOURce:]LIST:COUNt? [MIN|MAX,] (@<chanlist>)
Sets the list repeat count. This sets the number of times that a list is executed before it completes. The
count range is 1 through 4096. Use the Infinity parameter to run the list continuously.
Parameter Typical Return
1 4096|MIN|MAX|INFinity, *RST 1 <list count>
Sets the listcountto 10: LIST:COUN10, (@1)
l Use ABORt:TRANsient to stop the list at any time. When the list is aborted, the output returns to
the settings that were in effect before the list started.
l The query returns the list count for each channel specified. If a repeat count of 9.9E37 is returned,
it means the list is set to repeat continuously.
[SOURce:]LIST:CURRent[:LEVel] <value>{,<value>}, (@<chanlist>)
[SOURce:]LIST:CURRent[:LEVel]? (@<chanlist>)
[SOURce:]LIST:VOLTage[:LEVel] <value>,{<value>}, (@<chanlist>)
[SOURce:]LIST:VOLTage[:LEVel]? (@<chanlist>)
Specifies the setting for each list step. Values are specified in either amperes or volts. A comma-
delimited list of up to 512 steps may be programmed.
Parameter Typical Return
0 to maximum. The maximum valuedepends on rating of the power mod-
ule.
*RST 1 step set to the minimum programmable value.
<list value 1>,
<list value 2>,
<list value 3>...
Programs a current listof 3 steps. LIST:CURR 3,2,1, (@1)
Programs a voltagelistof 3 steps. LIST:VOLT 20,10,5, (@1)
l The order in which the list values are entered determines the sequence when the list executes.
l To create a valid list, the Voltage, Current, BOST, EOST, and Dwell lists must all be the same
length or have a length of 1, which is interpreted as having the same length as the longest list.
l This command overwrites any previously programmed list; it does not append to the previous list.