Sense Commands
Sense commands control the measurement ranges, datalog, Elog, histogram, sweep, and window
functions.
SENSe:CURRent:CCOMpensate 0|OFF|1|ON, (@<chanlist>)
SENSe:CURRent:CCOMpensate? (@<chanlist>)
This command does NOT apply to models .
On models N676xA this command only applies in the High current range.
Enables or disables the capacitive current compensation.
When turned ON, this function corrects the output current measurement for the current that flows into
the output capacitor. This current is only significant when high dv/dt occurs, and because this current
does not flow into the output load, it represents an error when not corrected. Note that turning on the
current measurement compensation increases the peak to peak noise in current measurements.
Parameter Typical Return
0|OFF|1|ON, *RST ON 0or1
Disablesense current compensation: CURR:CCOM OFF, (@1)
SENSe:CURRent[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects a DC current measurement range. The value that you enter must be the highest value that
you expect to measure.Units are in amperes. The instrument selects the range with the best
resolution for the value entered.
Parameter Typical Return
0 to maximum|MIN|MAX, *RST MAX <max current>
Selects the 2 A range: SENS:CURR:RANG 2, (@1)
l When queried, the returned value is the maximum DC current that can be measured on the range
that is presently set.
SENSe:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:CURRent[:DC]:RANGe:AUTO? (@<chanlist>)
Enables/disables seamless measurement autoranging.