Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. See the Syntax Conventions for SCPI.
ABORt
:ACQuire (@chanlist) Cancels any triggered measurements.
:DLOG Stops internal data logging.
:ELOG (@chanlist) Stops external data logging.
:HISTogram (@chanlist) Stops the histogram current measurement. (only on N6781A/82A/85A/86A)
:TRANsient (@chanlist) Cancels any triggered actions.
CALibrate
:COUNt? Returns the number of times the unit has been calibrated.
:CURRent
[:LEVel] <value>, (@channel) Calibrates the current programming.
:LIMit
:NEGative <value>, (@channel) Calibrates negative current limit. (only on N678xA SMU, N6783A)
:POSitive <value>, (@channel) Calibrates positive current limit. (only on N678xA SMU)
:MEASure <value>, (@channel) Calibrates the current measurement.
:PEAK (@channel) Calibrates peak current limit (only on N675xA, N676xA)
:DATA <value> Enters the calibration value read by the external meter.
:DATE <"date">, (@channel) Stores the calibration date in nonvolatile memory.
:DPRog (@channel) Calibrates the current downprogrammer.
:LEVel P1|P2|P3 Advances to the next level in the calibration.
:PASSword <value> Sets a numeric password to prevent unauthorized calibration.
:RESistance 20 | 6, (@channel) Calibrates the output resistance. (only on N6781A, N6785A)
:SAVE Saves the calibration constants in non-volatile memory.
:STATe 0|OFF|1|ON Enables or disables calibration mode.
:VOLTage
[:LEVel] <value>, (@channel) Calibrates the voltage programming.
:CMRR, (@channel) Calibrates the voltage common mode rejection ratio. (N675xA, N676xA)
:LIMit
:POSitive <value>, (@channel) Calibrates positive current limit. (only on N678xA SMU)
:MEASure <value>, (@channel) Calibrates the voltage measurement.
:AUXiliary, (@channel) Calibrates the auxiliary voltage measurement. (only on N6781A, N6785A)
DISPlay
[:WINDow]
[:STATe] 0|OFF|1|ON Turns the front panel display on or off.
:VIEW METER1 | METER4 Selects 1-channel or 4-channel meter view.
1 Quick Reference
26 Keysight N6705C Operating and Service Guide