Meter View
Press Meter
View
This key
toggles
between
multipleand
singleoutput
views.
Multiple Output View Single Output View
1. Output
Identifier
Identifies the output. When an outputis selected, the background becomes highlighted. The selected output is displayed in an
enlarged format in singleoutputview.
2. Output
Status
Off: the output is off
CV: the outputis in constantvoltage mode
CC: the outputis in constant current mode
Unr: the output is unregulated
CP+, CP–: a positive or negativepower limit
CL+, CL–: a positive or negative current limit
VL+, VL–: a positive or negative voltage limit
OV: over-voltage protection tripped
OV–: negative voltage protection tripped
OC: over-current protection tripped
OT: over-temperatureprotection tripped
PF: a power-fail condition occurred
Inh: an external inhibit signal is received
Osc: the oscillation protection tripped
Prot: a coupled outputprotect occurred
3. Output
Meter
Displays the actual output voltage and current. Displays power in singleoutputview.
4. Output
Setting
Displays the present output voltageand current settings. Turn the front panel voltageor current knob to adjustthese settings.
Can also bechanged using the numeric keypad.
5. Remote
Status
Error an error has occurred (press the Menu key, select Utilities, then Error Log)
LAN the LANis connected and has been configured
IO there is activity on oneof the remoteinterfaces
6. Model
Number
Identifies the model number of the power moduleconnected to this output.
7. Arb, Delay,
& Slew Rate
Displays the Arb waveform that is presently configured for this output. If no Arb is configured, no waveform will be displayed.
Also displays the OutputOn and Output Off delay settings as well as the slewratesetting.
8. Polarity
Reverse
Indicates that the outputand sense polarities are reversed.
9. Ratings &
Protection
Displays the maximum voltageand current ratings of the output. Also displays the present over-voltage protection setting
and whether over-current protection is on or off.
10. Other
Outputs
Displays the actual voltage, current, and status of the other outputs.
Keysight N6705C Operating and Service Guide 19