select a maximum of five measurements to be displayed.
From the remote interface:
The following commands position the markers. To position the two data log markers at 100 seconds
and at 200 seconds from the start trigger on the data log:
SENS:DLOG:MARK1:POIN 100
SENS:DLOG:MARK2:POIN 200
The following commands return data between the two markers. To return the average current or
voltage between the markers:
FETC:DLOG:CURR? (@1)
FETC:DLOG:VOLT? (@1)
To return the minimum current or voltage between the markers:
FETC:DLOG:CURR:MIN? (@1)
FETC:DLOG:VOLT:MIN? (@1)
To return the maximum current or voltage between the markers:
FETC:DLOG:CURR:MAX? (@1)
FETC:DLOG:VOLT:MAX? (@1)
To return the peak-to-peak current or voltage between the markers:
FETC:DLOG:CURR:PTP? (@1)
FETC:DLOG:VOLT:PTP? (@1)
Data Logger Presets
Select the Preset button to return the Data Logger View to the power-on display settings. The vertical
offset of each trace will be set to a different value. This is to prevent the traces from overlapping. The
offset is referenced to the horizontal center line of the grid.
Keysight N6705C Operating and Service Guide 157
4 Using the Measurement Functions