Parameter Typical Return
(none) <value> [,<value>]
or <Block>
Returns the measured currentarray FETC:ARR:CURR? (@1)
Returns the measured voltage array FETC:ARR:VOLT? (@1)
Returns the measured power array FETC:ARR:POW? (@1)
FETCh:DLOG:AHOur? (@<chanlist>)
FETCh:DLOG:WHOur? (@<chanlist>)
Returns the amp-hours or watt-hours delineated by the front panel markers on the selected data log
traces that appears in the instrument's Data Logger view. These values corresponds to the Ah (amp-
hours) or Wh (watt-hour) value shown on the bottom of the instrument's Data Logger view when
markers are displayed. The reading is in the form +n.nnnnnnE+nn for each channel specified. Multiple
responses are separated by commas.
Parameter Typical Return
(none) <amp-hour value>
<watt-hour value>
Returns the amp-hours delineated by the markers FETC:DLOG:AHO? (@1)
Returns the watt-hours delineated by the markers FETC:DLOG:WHO? (@1)
FETCh:DLOG:CURRent[:DC]? (@<chanlist>)
FETCh:DLOG:VOLTage[:DC]? (@<chanlist>)
Returns the averaged measurement.Values returned are either in amperes or volts. The reading is in
the form +n.nnnnnnE+nn for each channel specified. Multiple responses are separated by commas.
Parameter Typical Return
(none) <datalog value>
Returns the averaged current FETC:DLOG:CURR? (@1)
Returns the averaged voltageFETC:DLOG:VOLT? (@1)
FETCh:DLOG:CURRent:MAXimum? (@<chanlist>)
FETCh:DLOG:VOLTage:MAXimum? (@<chanlist>)
FETCh:DLOG:CURRent:MINimum? (@<chanlist>)
FETCh:DLOG:VOLTage:MINimum? (@<chanlist>)
Returns the maximum or minimum value. Values returned are either in amperes, or volts. See
Measurement Data Formats.The reading is in the form +n.nnnnnnE+nn for each channel specified.
Multiple responses are separated by commas.
Keysight N6705C Operating and Service Guide 283
7 SCPI Programming Reference