INITiate:CONTinuous:TRANsient 0|OFF|1|ON, (@<chanlist>)
INITiate:CONTinuous:TRANsient? (@<chanlist>)
Continuously initiates the transient trigger system. This allows multiple triggers to generate multiple
output transients.
Parameter Typical Return
0|OFF|1|ON, *RST ON 0or1
Continually initiates the trigger system: INIT:CONT:TRANON, (@1)
l With continuous triggering disabled, the trigger system must be initiated for each trigger using the
INITiate:TRANsient command.
l ABORt:TRANsient does not abort continuous triggers if INITiate:CONTinuous:TRANsient ON has
been programmed. In this case, the trigger system will automatically re-initiate.
Keysight N6705C Operating and Service Guide 295
7 SCPI Programming Reference