[SOURce:]ARB:CURRent:STAircase:END:TIMe <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END:TIMe<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END:TIMe? [MIN|MAX,] (@<chanlist>)
Specifies the time in seconds, after the staircase completes, that the ending current or voltage level
persists. Referenced to t2 in the Staircase diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time in seconds, after the staircase completes, that the ending current or voltage level
persists for Arb sequence steps.
Parameter Typical Return
0 – 262.144 | MIN | MAX, *RST 0 <end time>
Step 0 - 99, *RST 0 (none)
Programs an end time after the current staircase:ARB:CURR:STA:END:TIM 2,(@1)
Programs an end time after the voltagestaircase:ARB:VOLT:STA:END:TIM 2,(@1)
Programs an end time for step 0:ARB:SEQ:STEP:CURR:STA:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:STAircase:NSTeps <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:NSTeps<value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:NSTeps? [MIN|MAX,] (@<chanlist>)
Specifies the number of steps in the staircase.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:NSTeps <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:NSTeps? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:NSTeps <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:NSTeps? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the number of steps in the staircase for Arb sequence steps.
Parameter Typical Return
0 – 500 | MIN | MAX, *RST 10 <number of steps>
Step 0 - 99, *RST 0 (none)
Programs five steps for the current staircase: ARB:CURR:STA:NST5,(@1)
Programs five steps for the voltagestaircase: ARB:VOLT:STA:NST 5,(@1)
Programs five steps for sequence step 0: ARB:SEQ:STEP::CURR:STA:NST 5, 0(@1)
7 SCPI Programming Reference
250 Keysight N6705C Operating and Service Guide