Staircase
Back to Arb Commands
Arb Type Current Voltage
Staircase
[SOURce:]ARB:CURRent:STAircase:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STAircase:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STAircase:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the ending current or voltage level after the staircase occurs. Referenced to I1 and V1 in the
Staircase diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STAircase:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STAircase:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the ending current or voltage level after the staircase occurs for Arb sequence steps.
The difference between the start level and the end level is divided equally among the
number of steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
The maximum value varies, depending on the currentor voltage
rating of the power module.
<end value>
Step 0 - 99, *RST 0 (none)
Programs an end level current: ARB:CURR:STA:END 2,(@1)
Programs an end level voltage: ARB:VOLT:STA:END 20,(@1)
Programs an end level current for step 0: ARB:SEQ:STEP:CURR:STA:END 2, 0,(@1)
Keysight N6705C Operating and Service Guide 249
7 SCPI Programming Reference