Step
Back to Arb Commands
Arb Type Current Voltage
Step
[SOURce:]ARB:CURRent:STEP:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STEP:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:END[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:END[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the step current or voltage level. Referenced to I1 and V1 in the Step diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:STEP:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:STEP:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STEP:END[:LEVel] <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:STEP:END[:LEVel]? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the step current or voltage level for Arb sequence steps.
Parameter Typical Return
Current: 0 - maximum|MIN|MAX, *RST MIN
Voltage: 0 - maximum|MIN|MAX, *RST MIN
The maximum value varies, depending on the currentor voltage
rating of the power module.
<end value>
Step 0 - 99, *RST 0 (none)
Programs the step level current: ARB:CURR:STEP:END 2,(@1)
Programs the step level voltage: ARB:VOLT:STEP:END 20,(@1)
Programs a step level for step 0: ARB:SEQ:STEP:CURR:STEP:END 2, 0,(@1)
[SOURce:]ARB:CURRent:STEP:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:CURRent:STEP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:STARt[:LEVel] <value>, (@<chanlist>)
[SOURce:]ARB:VOLTage:STEP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the initial current or voltage level. Referenced to I0 and V0 in the Step diagrams.
Keysight N6705C Operating and Service Guide 253
7 SCPI Programming Reference

Table of Contents