Current Measure - the Current Measure selection emulates a zero-burden ammeter. Voltage priority
mode is set by default. The +/- Current Limits are set to their maximum values. The current
measurement cannot exceed the current ratings of the unit as indicated in the +/- Current Limit fields.
The following additional settings are also required. Remote sensing (4 Wire) is always required. For
applications with transient voltage drops, use the High3 setting with a compensation capacitor across
the DUT. Capacitors can be film or ceramic with values from 7 µF to 150 µF. Refer to Keysight N678xA
SMU Wiring for additional wiring information.
4 Using the Measurement Functions
130 Keysight N6705C Operating and Service Guide