1 parameter (voltage or current)
3 parameters (voltage+min+max)
6 parameters (voltage+min+max X 2 outputs)
12 parameters (voltage+min+max X 4 outputs)
24 parameters (voltage+min+max X 4 outputs
& current+min+max X 4 outputs)
20 microseconds (rounded)
60 microseconds (rounded)
120 microseconds (rounded)
240 microseconds (rounded)
480 microseconds (rounded)
When the Power trace is selected, it counts as 2 parameters, since voltage and current
must be measured to calculate power. If voltage and current traces have already been
selected, the Power trace is not counted as a parameter
Standard mode (interleaved)
Standard (interleaved) mode only applies when both voltage and current measurement traces are
selected on power modules other than Models N676xA and N678xA SMU. These other power modules
cannot measure voltage and current simultaneously; hence, the voltage and current measurements
must be interleaved for all power modules in the mainframe. Each measurement is sampled for about
5 milliseconds at the beginning of every sample period. Power is calculated from the interleaved
measurements.
Standard data sampling is used with the following Power Module and Display Trace selections:
Power Module Display Trace selection
N673xB, N674xB Voltage, Current, and Power
N675xA, N677xA Voltage, Current, and Power
N6783A-BAT/MFG Voltage, Current, and Power
Available capabilities for ALL outputs
• Sample period: 20.48 microseconds to 60 seconds
• Trigger source: all trigger sources available
• Trigger offset: 0 to 100%
• Values logged: average, minimum, maximum (minimum/maximum values must be selected)
Keysight N6705C Operating and Service Guide 159
4 Using the Measurement Functions