Parameter Typical Return
(none) <option number>
Returns installed options: *OPT?
*RCL <0-9>
Recall State Restores the instrument to a state that was previously stored in memory locations 0
through 9 with the *SAV command. All instrument states are recalled except: (1) the trigger system is
set to the Idle state, (2) calibration is disabled, (3) trigger settings are set to their Idle state, and (4)
non-volatile settings are not affected.
Parameter Typical Return
0 - 9 (none)
Recall state from location 1: *RCL 1
l Location 0 is automatically recalled at power turn-on when the Output Power-On state is set to
RCL0.
l You cannot recall an instrument state from an empty or deleted storage location.
l Stored instrument states are affected by *RST.
*RDT?
Returns the output channel descriptions. The description is an arbitrary ASCII string with up to four
semicolon-separated fields. To read the string into your computer, be sure to dimension a string
variable with at least 50 characters.
Parameter Typical Return
(none) <ASCII string with semicolon-separated fields>
Identify the mainframe's installed power modules: *RDT?
l Example 4-channel response: CHAN1,N6751A;CHAN2,N6752A;CHAN3,N6761A;CHAN4,N6762A
*RST
Reset Command Resets the instrument to pre-defined values that are either typical or safe. These
settings are described in Factory Reset State.
Parameter Typical Return
(none) (none)
Reset the instrument: *RST
l *RST forces the ABORt command. This cancels any trigger actions presently in process, and resets
the WTG bit in the Status Operation Condition register.
Keysight N6705C Operating and Service Guide 291
7 SCPI Programming Reference